Description

Certified reference material of agricultural soil. Silty clay loam heavily contaminated with some heavy metals.
qAN-ZP02
In stock
Contact us for a price
+
Features
Description
Documents
Product inquiry
Package:
glass
RM Matrix:
  • Soil
Type:
CRM
Weight:
50 g

AN-ZP02

Silty clay loam contaminated with some heavy metals

contents of element extractable fractions with aqua regia according to ISO 11466

Packing:

50 g of powder of the respective soil with a particle size of less than 100 µm in amber glass bottle.

Intended use:

This certified reference material is intended primarily for use in realizing metrological traceability, validation of analytical methods and the transfer of property values by total element contents to other materials of similar composition.

Recommended analytical methods for method dependent measurements:

Flame AAS, ETA-AAS, ICP-OES, ICP-MS, AMA-254

Certified and indicative values of contents of element fractions extractable by aqua regia in CRM AN-ZP02 and their associated uncertainties.

Element Concentration [mg/kg] Uncertainty [mg/kg]
As 16.0 1.0
Ba 158 7
Be 0.64 0.05
Cd 1.63 0.11
Co 27.2 1.1
Cr 68.1 2.3
Cu 22.0 1.2
Hg-total 0.136 0.010
Hg-extract (0.149) -
Mn 1 802 70
Mo 0.75 0.07
Ni 9.75 0.57/td>
Pb 205 8
Sb 8.03 0.54
Sn (4.65) -
Sr 31.6 1.5
Tl N.M. -
V 138 3
Zn 173 7

All values corrected to a dry weight at 105 °C
Uncertainty - expanded combined uncertainty (k=2)
() Indicative value
N.M. not measured

Matrix composition of CRM AN-ZP02 (% of the oxides dry weight at 105°C)

Analyte Matrix composition (% oxides, dry weight)
SiO2 60.30
Al2O3 12.92
CaO 2.07
MgO 1.62
MnO 0.263
Li2O 0.006
Fe2O3 8.33
K2O 0.99
Na2O 2.89
P2O5 0.212
TiO2 1.56
Total nitrogen 0.17
H2O (105 °C) 2.61
Loss on ignition (900 °C) 6.67
Certificate COA-ANZP01 ANZP02-EN-V02 (COA-ANZP01 ANZP02-EN-V02.pdf, 726 Kb) Download